High temporal resolution coordinate-sensitive detector with gallium-arsenide photocathode

V. G. Debur, G. M. Beskin, S. V. Karpov, V. L. Plokhotnichenko, A. S. Terekhov, S. S. Kosolobov, G. E. Shaibler

Результат исследований: Вклад в журналСтатьярецензирование

6 Цитирования (Scopus)

Аннотация

We present a coordinate-sensitive detector with a gallium-arsenide photocathode-a vacuum photoelectronic device featuring microchannel gain and a multi-element collector-which had several prototypes manufactured. Quantum efficiency of the photocathode amounts to 48% at maximum, its sensitivity range spans from 350 to 900 nm. The microchannel gain unit consists of two plates with an ion barrier film at the entrance. The coordinates of electron avalanches are determined using a 16-element collector. The spatial resolution, diameter of the working field, and temporal resolution of the detector are equal to 25 μm, 18 mm, and about 1 μs, respectively.

Язык оригиналаАнглийский
Страницы (с-по)386-391
Число страниц6
ЖурналAstrophysical Bulletin
Том64
Номер выпуска4
DOI
СостояниеОпубликовано - окт. 2009
Опубликовано для внешнего пользованияДа

Fingerprint

Подробные сведения о темах исследования «High temporal resolution coordinate-sensitive detector with gallium-arsenide photocathode». Вместе они формируют уникальный семантический отпечаток (fingerprint).

Цитировать